Temperature Surface Measurements With A 3d Velocity Fluid Flow Measurements System Using The Same Laser Source And A Single Instrument Artículo académico uri icon

Abstracto

  • The combination of flow velocimetry techniques and Temperature Sensitive Paints, (TSP), requires working with different laser beam intensities. Because velocity flow measurements (i.e. Particle Image Velocimetry, PIV) needs high level laser power compared with temperature surface measurement, where lower levels of laser power is required, is necessary to adjust the system to avoid the damage of the paint due to the high intensities in laser velocimetry measurements. The use of a paint of different grey levels, from white to black, as backgrounds above the TSP film deposition allows to make both, velocity and temperature measurements with the same laser power without damaging the TSP. This work is centered in the characterization, testing and calibration improvements of the temperature surface measurements using Temperature Sensitive Paints as a part of the 3D tunneling velocimetry system.

fecha de publicación

  • 2005

Volumen

  • 5776